3.5 Electrode installation and measuring cable connection
Measuring cable connection
The terminal block is located in a separate
connection compartment which is accessible
by opening the plastic cover. Remove the
pre-pressed knock-outs for cable entry.
The pH and redox electrodes are connected
using special terminated and shielded multi-
core cables. The measuring cable can be
extended with junction box VBA or VBM.
Termination instructions are supplied with the
measuring cables.
Note:
Protect connectors, cable ends
and terminals against moisture to
prevent inaccurate measurement!
For futher information on cables
and junction boxes refer to chap.
9 Accessories.
Structure and termination of the measuring cable
Measuring cable requirements
Sensor type Cable Extension
Electrode without temperature sensor CPK 1 VBA / VBM box + CYK 71 cable
Electrode with temperature sensor Pt 100 and
TOP 68 plug-in head
CPK 9 VBA / VBM box + CYK 71 cable
IsFET electrode with temperature sensor
Pt 100 / Pt 1000 and TOP 68 plug-in head
CPK 12 VBA / VBM box + CYK 12 cable
pH single electrode with reference electrode
and separate temperature sensor
CPK 2 VBA / VBM box+ PMK cable
CPK1_12e.EPS
Fig. 3.13
Construction of special
measuring cable
Installation Liquisys M CPM 223 / 253